Change of https://www.gov.uk/guidance/uk-standards-for-air-quality-monitoring-instruments

Change description : 2026-09-25 16:00:00: First published. [Guidance and regulation]

Showing diff : ..2026-09-25 15:00:35.342436902 +00:00

Guidance

UK standards for air quality monitoring instruments

The standard instruments, methods and requirements for monitoring air pollution in the UK.

MCERTS performance standards 

Monitoring instruments that are used in UK air pollution monitoring networks must pass the tests and requirements in the MCERTS performance standards. 

The MCERTS performance standards make sure that the instruments used to monitor air pollution comply with the: 

  • Air Quality Standards Regulations 2010 
  • Environmental Targets (Fine Particulate Matter) (England) Regulations 2023 (England only) 

Instrument standards for the Automatic Urban and Rural Network (AURN) 

The following table shows the instrument performance standards for common air pollutants. Instruments must meet these standards before they can be approved and certified for use in the AURN. 

The documents are produced by the European Committee for Standardisation and published by the British Standard Institute. 

Pollutant Performance standard Details
Benzene EN 14662-1:2023 Standard method for measuring benzene by pumped sampling, thermal desorption and capillary gas chromatography
Carbon monoxide (CO) EN 14626:2024 Standard method for measuring carbon monoxide by non-dispersive infrared spectroscopy
Nitrogen oxides (NO and NO2) BS EN 14211:2024 Standard method for measuring nitrogen dioxide and nitrogen monoxide by chemiluminescence
Ozone (O3) BS EN 14625:2024 Standard method for measuring ozone by ultraviolet photometry
PM2.5 and PM10 (gravimetric) BS EN 12341:2023 Standard gravimetric method for measuring PM2.5 and PM10 in suspended particulate matter
PM2.5 and PM10 (continuous) BS EN 16450:2017 Automated measuring systems for measuring the concentration of particulate matter (PM2.5 and PM10)
Sulphur dioxide (SO2) BS EN 14212:2024 Standard method for measuring sulphur dioxide by ultraviolet fluorescence

Monitoring instruments that measure these pollutants in the AURN must get ‘type approval’ to show they have met the performance standard. Type approval is granted by CSA Group Ltd, the certification body appointed by the Environment Agency. 

Testing AURN monitoring instruments against these standards began in 2005. 

See a list of monitoring instruments certified by MCERTS. 

Find out how the AURN monitors air pollution. 

Instrument standards for measuring PM2.5 and PM10 

PM2.5 is particulate matter less than 2.5 micrometres in diameter. PM10 is particulate matter less than 10 micrometres in diameter. 

The standard methods for measuring PM2.5 and PM10 are: 

  • BS EN 16450:2017 ambient air (automated system for measuring the concentration of PM2.5 and PM10) 
  • BS EN 12341:2023 ambient air (standard gravimetric method for measuring PM2.5 and PM10 in suspended particulate matter – the reference method) 

Real-time, automatic monitoring 

BS EN 16450:2017 sets the minimum performance standard and test procedure for automated continuous measurement systems for PM2.5 and PM10. These systems provide data in near real-time. 

Before an automatic PM2.5 or PM10 instrument can be included in the AURN, it must have certification to show that it is equivalent to the standard gravimetric reference method (BS EN 12341:2023). 

To demonstrate this equivalence, follow the methods in the guide to the demonstration of equivalence of ambient air monitoring. 

Find more information about PM2.5 and PM10 equivalence in the:  

Non-real-time, non-automatic monitoring 

BS EN 12341:2023 sets the standard operating method for gravimetric samplers. However, the filters in gravimetric samplers must be weighed in a laboratory before and after exposure. Because this takes time, the data is not real-time, hourly data.  

Additional standards for measuring PM2.5 

The Environment Agency requires additional MCERTS certification for measuring PM2.5. This makes sure that all instruments demonstrate equivalence in a representative particulate matter pollution climate for the UK. 

To get this certification, monitoring instruments must meet the MCERTS performance standards. 

List of approved particulate matter monitoring instruments 

This list is in alphabetical order. The list will be updated when new instruments are approved. 

Manufacturer and instrument name PM size fraction Correction required
Envea MP101M PM2.5 None
Envea MP101M PM10 None
European reference method in compliance with EN 12341:2023 PM10 None
European Reference Method in compliance with EN 12341:2023 PM2.5 None
FAI SWAM 5a 1-Hour dual channel PM2.5 and PM10 in the same instrument None
FAI SWAM 5a 24-hour dual channel PM2.5 and PM10 in the same instrument None
FAI SWAM 5a 24-hour single channel PM10 None
FAI SWAM 5a 24-hour single channel PM2.5 None
Grimm EDM280 PM10 None
Grimm EDM280 PM2.5 Subtract 1.032 and then divide by 1.022
Met One Smart Heated BAM 1020 PM10 Divide by 1.035
Met One Smart Heated BAM 1020 PM2.5 None
Met One Unheated BAM 1020 PM10 Multiply by 0.833
Opsis SM200 PM10 None required for real time beta attenuation results. If filter masses are reported, subtract 1.286 then divide by 0.819
Palas Fidas 200, 200E or 200S operating with algorithm method 11’ PM2.5 and PM10 in the same instrument None required for PM10. Divide by 1.06 for PM2.5
Palas Fidas 200, 200E or 200S operating with algorithm method 11’ PM2.5 only Divide by 1.06 for PM2.5
Thermo Fisher FDMS 1405DF PM2.5 and PM10 None
Thermo Fisher FDMS 1405F PM10 None
Thermo Fisher FDMS 1405F PM2.5 None
Thermo Fisher FDMS 8500 B, C or CB PM10 None
Thermo Fisher FDMS 8500 B, C or CB PM2.5 None
Thermo Fisher Partisol 2025 A, B or i PM10 None
Thermo Fisher Partisol 2025 A, B or i PM2.5 None

Future changes to the MCERTS standards   

Instrument standards will be updated in the future to adapt to the changing levels and composition of PM2.5. 

Until the standards are reviewed, monitoring instruments must meet the current MCERTS standards. 

MCERTS standards under review 

The following standards are currently being reviewed and redeveloped: 

Defra will update this page when the new standards are published.

Updates to this page

Published 25 September 2026

Sign up for emails or print this page

Update history

2026-09-25 16:00